Edit inspection #5 no serial number

Ready

Head Information

1. Internal Head Sensor — single measurement per head

Parameter Standard Reading Result
Contact Detection — Pressure 70 ± 4 kPa pending
Internal Head Sensor — Flow Rate ≥ 3.5 L/min pending

2. Readings per shaft (A–T)

Entry runs column by column: Enter steps from A through T within one column, then jumps to the top of the next. Vacuum Pressure accepts values with or without a minus sign (−88 and 88 are treated alike); the decimal separator may be a point or a comma. Internal Head Sensor Pressure: shaft A is the reference shaft, calibrated to 65 kPa (tolerance ± 2.5 kPa). All other shafts are acceptable up to ± 10 kPa; beyond ± 5 kPa deviation the value is flagged as borderline in yellow.

Shaft Internal Head Sensor Vacuum — Z1 Vacuum — Z2 Vacuum Break Down Nozzle Cleaning Valve Air Stick Overall
Pressure (kPa)
A: 65 ± 2.5 · B–T: 65 ± 10
Pressure (kPa)
|Wert| ≥ 80
Flow (L/min)
≥ 2.0
Pressure (kPa)
|Wert| ≥ 80
Flow (L/min)
≥ 2.0
Pressure (kPa)
10 – 15
Flow (L/min)
≥ 0.5
Pressure (kPa)
≥ 100
Flow (L/min)
≥ 0.8
Repeated Sliding (mm)
|Wert| ≤ 0.08
A —
B —
C —
D —
E —
F —
G —
H —
I —
J —
K —
L —
M —
N —
O —
P —
Q —
R —
S —
T —
Within specification Borderline — acceptable but conspicuous Out of specification

3. Summary

INCOMPLETE20 of 20 shafts not yet fully recorded
  • 0 PASS
  • 0 borderline
  • 0 FAIL
  • 20 pending
20 shafts in total (A–T)